• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS HIGH SPEED ELEC G1 HANDPIECE AIR COOLED; MOTOR, DRILL, ELECTRIC - HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS HIGH SPEED ELEC G1 HANDPIECE AIR COOLED; MOTOR, DRILL, ELECTRIC - HANDPIECE Back to Search Results
Catalog Number EG1A_LL
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/04/2016
Event Type  malfunction  
Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during a lumbar laminectomy surgical procedure, it was discovered that once the foot pedal device was pressed, the burr device slowly came out of the hand piece device.According to the reporter, the burr device was secured in the hand piece device.It was reported that the user pulled on the cutting tip to ensure that the burr device was secured.The reporter indicated that when the hand piece device became operational, the cutting burr device migrated out of the drill.There were no delays to the surgical procedure as a spare device was available to complete the surgery.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device.A functional assessment was performed.The device was tested with various cutter devices and it passed all manufacturing specifications.The reported condition was not duplicated or confirmed.Therefore, an assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH SPEED ELEC G1 HANDPIECE AIR COOLED
Type of Device
MOTOR, DRILL, ELECTRIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5657466
MDR Text Key46470885
Report Number1045834-2016-11727
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK133604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEG1A_LL
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-