DEPUY SYNTHES POWER TOOLS HIGH SPEED ELEC G1 HANDPIECE AIR COOLED; MOTOR, DRILL, ELECTRIC - HANDPIECE
|
Back to Search Results |
|
Catalog Number EG1A_LL |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/04/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during a lumbar laminectomy surgical procedure, it was discovered that once the foot pedal device was pressed, the burr device slowly came out of the hand piece device.According to the reporter, the burr device was secured in the hand piece device.It was reported that the user pulled on the cutting tip to ensure that the burr device was secured.The reporter indicated that when the hand piece device became operational, the cutting burr device migrated out of the drill.There were no delays to the surgical procedure as a spare device was available to complete the surgery.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device.A functional assessment was performed.The device was tested with various cutter devices and it passed all manufacturing specifications.The reported condition was not duplicated or confirmed.Therefore, an assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|