MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-723RNAB
Device Problem Intermittent Continuity (1121)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/10/2016
Event Type  malfunction  
Manufacturer Narrative
Currently, it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call that he has to rewind the insulin pump multiple times to get the piston all the way fully rewound.Customer also mentioned that when doing a bolus it sometimes stops the bolus delivery.Customer stated that there were no alarms occurring.Customer was advised to call back if the issue persists.Insulin pump is being returned for analysis.
 
Manufacturer Narrative
The insulin pump passed all functional testing including the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.The insulin pump rewinds properly.The insulin pump was monitored in 50c oven for several days with a 1.0u basal rate and bolus deliveries and all registered properly in the daily totals history noted.The motor passed motor test.However, the insulin pump was received with cracked case at the display window corner, cracked reservoir tube lip, cracked battery tube threads and minor scratched lcd window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5657683
MDR Text Key45419414
Report Number2032227-2016-10779
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723RNAB
Device Catalogue NumberMMT-723RNAB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received05/10/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age78 YR
Patient Weight100
-
-