Model Number 0.850.0009
Device Problem Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/31/2016
Event Type  malfunction  
Manufacturer Narrative
The complete device was not returned however the battery handset and charging cradle were returned and are currently undergoing investigation.A follow-up report will be submitted upon the investigation completion.
 
Event Description
It was reported that the nomad pro battery handset started to emit smoke while on the charging cradle.There were no injuries reported.
 
Manufacturer Narrative
The analysis of the returned device has been completed.Our records indicate that the device was assembled in march 2011.The returned unit was inspected visually.The investigation found evidence of heavily applied dental office cleaners and fluid ingress into the interior of the handset.It was determined that there was a short circuit in the lower battery pack circuit board that lead to the battery failure.No injuries were reported.The nomad pro operator manual provides instructions on the proper care, cleaning, and maintenance of the device.This concludes our investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key5659687
MDR Text Key46013755
Report Number1017522-2016-00010
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date 03/31/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0009
Device Catalogue Number0.850.0009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/31/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-