It was reported by a company representative that her usb serial data cable was very loose, and requested a replacement cable.The usb serial data cable was returned for analysis which was completed 04/20/2016.During the analysis, it was identified that a known good tablet was unable to establish communication with a known good generator using the returned cable.The cause for the anomaly was associated with two disconnected wire connections in the returned serial cable.Once the wires were soldered onto the serial cable printed circuit board, no further anomalies were identified.No additional relevant information has been received to-date.
|