• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS - HOUSTON PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Disconnection (1171); Loose or Intermittent Connection (1371); Mechanical Problem (1384); Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 04/20/2016
Event Type  malfunction  
Event Description
It was reported by a company representative that her usb serial data cable was very loose, and requested a replacement cable.The usb serial data cable was returned for analysis which was completed 04/20/2016.During the analysis, it was identified that a known good tablet was unable to establish communication with a known good generator using the returned cable.The cause for the anomaly was associated with two disconnected wire connections in the returned serial cable.Once the wires were soldered onto the serial cable printed circuit board, no further anomalies were identified.No additional relevant information has been received to-date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5660109
MDR Text Key46585738
Report Number1644487-2016-01044
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2016
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received04/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/21/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-