MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00027-120
Device Problem Failure to Fire (2610)
Patient Problem No Patient Involvement (2645)
Event Date 04/26/2016
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The rep found no power to the imaging system.A check of the fuses f11 and 12 found that they were blown.The rep replaced the fuses.The imaging system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation as the suspect fuses were discarded at the site.
 
Event Description
A site representative reported that the imaging system had stopped working.There was no patient involved with this concern.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
amos jarrette
826 coal creek circle
louisville, CO 80027
7208902082
MDR Report Key5673780
MDR Text Key45747769
Report Number1723170-2016-00923
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Biomedical Engineer
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/21/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-