• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SCD EXPRESS CONTROLLER; SCD CONTROLLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SCD EXPRESS CONTROLLER; SCD CONTROLLER Back to Search Results
Model Number U9525
Device Problems Break (1069); Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/23/2016
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon completion, a full detailed report will be provided.
 
Event Description
On (b)(6) 2016 the customer states the unit is damaged.Upon triage on (b)(6) 2016 the service tech found the unit had a damaged power cord with exposed copper wires.
 
Manufacturer Narrative
A review of the information in the complaint file indicates this investigation was performed by a medtronic technical center for the reported condition of; the unit is damaged.During triage, factory service found the unit had a damaged power cord with exposed copper wires.Therefore, this report will be based on information provided by the technical center.The scd express was evaluated and the customer reported issue of damaged along with factory service finding of damaged power cord with exposed copper wire were both confirmed; problem was isolated to damaged power cord.Visual examination revealed damaged power cord with exposed copper wire.The cause of the reported condition for the damaged power cord is due to customer related accidental damage.The damaged power cord will be scrapped to correct the reported issue.The unit passed all initial testing after failure analysis repair instructions were completed.Scd express was manufactured in 2006.A review of the device history record shows this device was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.Based on the investigation, no corrective action is needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCD EXPRESS CONTROLLER
Type of Device
SCD CONTROLLER
Manufacturer (Section D)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5675481
MDR Text Key45772546
Report Number3008361498-2016-00019
Device Sequence Number1
Product Code JOW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberU9525
Device Catalogue NumberU9525
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received06/03/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-