• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; RESPIRATORY GAS HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; RESPIRATORY GAS HUMIDIFIER Back to Search Results
Catalog Number 425-00
Device Problem Failure to Sense (1559)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/06/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product usage when the alleged issue was detected is unknown.The device was received by the manufacturer for evaluation, but the results of the investigation are incomplete at the time of this report.The device history record investigation did not show issues related to complaint.
 
Event Description
The customer alleges that the unit will not recognize the probe.
 
Manufacturer Narrative
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit failed the initial power connect test.When the power on button is released the unit immediately drops out with no response.This failure is indicative of a defective power supply pcb.The on-board power supply pcb was by-passed with a known good power supply pcb and the test was attempted again, this time successfully.Based on the investigation performed, the reported complaint was confirmed.It was determined that the power supply pcb is defective.All units being returned for service will have power supply pcbs replaced.This unit was manufactured 18 apr 2007 and will be replaced.
 
Event Description
The customer alleges that the unit will not recognize the probe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
RESPIRATORY GAS HUMIDIFIER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5675533
MDR Text Key45762508
Report Number3003898360-2016-00544
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received05/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-