MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA PRESSURE WIRE AERIS G8; TRANSDUCER, PRESSURE,CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA PRESSURE WIRE AERIS G8; TRANSDUCER, PRESSURE,CATHETER TIP Back to Search Results
Model Number C12058
Device Problems Sticking (1597); Improper or Incorrect Procedure or Method (2017); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/01/2016
Event Type  Injury  
Event Description
The pressurewire aeris became stuck in a stent.The physician pulled on the pressurewire with excessive force to withdraw it, causing the stent to become distorted.A stent was placed over the existing stent.The pressurewire distal tip was noted to be extended.No consequences were reported for the patient.
 
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.A review of the device history record was not possible since the batch number was unavailable.Based on the information received, the cause of the reported incident could not be conclusively determined.The pressurewire instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire instructions for use (ifu) states that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.The pressurewire instructions for use (ifu) states that the user should advance or withdraw the pressurewire slowly and never push, withdraw or torque the pressurewire if it meets resistance.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Our investigation was limited to the review of the device history record, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with sjm specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.The pressurewire instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire instructions for use (ifu) states that torquing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.The pressurewire instructions for use (ifu) states that the user should advance or withdraw the pressurewire slowly and never push, withdraw or torque the pressurewire if it meets resistance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE WIRE AERIS G8
Type of Device
TRANSDUCER, PRESSURE,CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key5677020
MDR Text Key45852327
Report Number3008452825-2016-00068
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K131452
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 05/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2018
Device Model NumberC12058
Device Catalogue NumberC12058
Device Lot NumberC5375742
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-