Catalog Number 500101
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reported that there is a small leak on the tah-t left cannula.The customer also reported that the 70cc tah-t cannula was repaired.There was no reported adverse patient impact as a result of the cannula leak and subsequent repair.This alleged failure mode poses a low risk to the patient because although there was a cannula leak, the tah-t continued to perform its life-sustaining functions.Syncardia has initiated a capa (corrective or preventive action) to address the cannula tear/leak issue.The capa will document the investigation including, but not limited to, potential root cause and corrective actions associated with all cannula tear events.The cannula piece will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The patient was implanted with the syncardia 70cc temporary total artificial heart (tah-t) on (b)(6) 2014 and had been supported by the tah-t system for 609 days when the customer reported that there was a small air leak inn the patient's tah-t left cannula.The customer also reported that the cannula was repaired by cutting off the section of the cannula with the air leak and reinserting the cpc connector.There was no reported adverse patient impact as a result of the cannula leak and subsequent repair.The damaged section of the 70cc tah-t cannula was not returned by the customer to syncardia, and therefore it could not be evaluated.Syncardia conducted a review of the tah-t device history record (dhr) and sterilization work order and confirmed that all manufacturing was performed to specifications, and the 70cc tah-t met all specified requirements prior to shipment.The patient was successfully transplanted on (b)(6) 2016.Syncardia has initiated a capa (corrective/preventive action) to address the cannula tear issue.The capa will document the investigation including, but not limited to, potential root cause and corrective actions associated with all cannula tear events.The syncardia freedom driver system guidebook for patients and caregivers - us, instructs the patients and caregivers to inspect the drivelines and cannulae daily.If a hole is observed in the drivelines or cannulae, household tape must be applied to temporarily repair the hole and the hospital contact person must be contacted to arrange for a hospital-based evaluation.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T)
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key5677195
MDR Text Key46473832
Report Number3003761017-2016-00209
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 05/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2017
Device Catalogue Number500101
Device Lot Number095781-SYNHALL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age25 YR
-
-