• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - SINGAPORE HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SINGAPORE HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310
Device Problem Filling Problem (1233)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/10/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the homechoice device overfilled a patient while performing peritoneal dialysis.The patient stated that the device would complete therapy in four cycles instead of five and they had no symptoms of an increased intraperitoneal volume event.The technical services representative (tsr) arranged to swap the device and discussed the use of manual supplies to complete therapy until the new device arrives.There was no patient injury or medical intervention reported.No additional information is available.
 
Manufacturer Narrative
(b)(4).The homechoice device was returned and evaluated by the product analysis lab (pal).A review of the device logs did not verify the reported issue.A device history review revealed no issues that could have caused or contributed to the reported issue.An internal/external inspection was performed and the device passed.A short simulated therapy was performed and passed.A check of the pneumatic system found the pneumatic system working to specifications.The device passed both the homechoice return instrument test/evaluation (rite) electrical test and the rite functional test.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park
singapore 2573
SN  2573
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park
singapore 2573
SN   2573
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5678185
MDR Text Key45862515
Report Number1416980-2016-09700
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C8310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age53 YR
-
-