MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; RESPIRATORY GAS HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; RESPIRATORY GAS HUMIDIFIER Back to Search Results
Catalog Number 425-00
Device Problem Failure to Sense (1559)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.The device history record investigation did not show issues related to complaint.A document assessment (fmea) was conducted and no changes required.Corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation and confirm the alleged defect.At this time since the sample is not available, it is not possible to determine the source of the defect reported.Customer complaint cannot be confirmed since the device sample is not available to perform a proper investigation and determine the root cause.If the device sample becomes available at a later date, this complaint will be updated accordingly.Teleflex will continue to monitor and trend on similar complaints.
 
Event Description
The customer alleges that the device is not registering a connection to the patient.No patient injury reported.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
 
Event Description
The customer alleges that the device is not registering a connection to the patient.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
RESPIRATORY GAS HUMIDIFIER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5681828
MDR Text Key45992405
Report Number3003898360-2016-00583
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-