TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1
|
Back to Search Results |
|
Model Number 801041 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/06/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The returned large roller pump did not boot up when power supply of system-1 was on.After they changed the returned cable to the other one, the large roller pump booted up.When they connected the returned cable to the other large roller pump, after a few minutes, the pump booted up.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the large roller pump did not boot-up.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
No part will be returned to the manufacturer for further evaluation.
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.Per manufacturer's subsidiary engineer, the unit has already been returned to customer.No part will be returned to the manufacturer for further evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|