Catalog Number TDVI-25-240
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use state: "if package is opened or damaged when received, do not use.Visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." the instructions for use also state to "disengage luer slip fitting from adjustment wheel prior to adjusting needle extension.Caution: attempting to adjust needle while luer slip fitting is engaged may result in damage to device." additionally, the instructions for use state to "verify adjustment by advancing and engaging luer slip fitting.Needle should move in and out of outer sheath easily.Inspect needle for breaks.Caution: advancing and retracting needle while catheter is coiled may result in stretching of catheter and damage to device." prior to distribution, all disposable varices injectors are subjected to a functional test to ensure appropriate needle extension.During packaging a visual inspection of the distal end of the sheath is conducted to insure the needle is not extending from the sheath.A review of the device history record confirmed that this lot met manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an isolated occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic procedure, the physician used a cook disposable varices injector.The catheter separated [unknown location] and did not pull back into the needle correctly.It's unknown how the procedure was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE VARICES INJECTOR
Type of Device
FBK, ENDOSCOPIC INJECTION NEEDLE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key5700586
MDR Text Key47843153
Report Number1037905-2016-00167
Device Sequence Number1
Product Code FBK
UDI-Device Identifier00827002218598
UDI-Public(01)00827002218598(17)190309(10)W3693234
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K851952
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 05/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberTDVI-25-240
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Device Age2 MO
Event Location Hospital
Date Manufacturer Received05/12/2016
Date Device Manufactured03/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No