CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/10/2016
Event Type  malfunction  
Event Description
It was reported that the patient could not have her vns system interrogated or adjusted due to an issue with the physician's programming system.The computer would reportedly not turn on.Attempts for additional pertinent information have been unsuccessful to date.
 
Event Description
Communication was received from the treating physician that confirmed the report of the failure of his programming system to power on.The identity of the product was also indicated at that time.No additional pertinent information has been received to date.
 
Event Description
Follow up with the physician's office allowed for troubleshooting.It was discovered that the 9v wand battery was depleted.After replacing the battery, the programming system functioned as expected by interrogated a demo generator.No additional pertinent information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5702067
MDR Text Key47896111
Report Number1644487-2016-01264
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Patient Family Member or Friend
Type of Report Initial,Followup,Followup
Report Date 05/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number521392
Other Device ID NumberVERSION 7.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received08/23/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age57 YR
-
-