There is evidence of benzalkonium interference routinely throughout the use life of the cartridge installed on sn (b)(4).Benzalkonium is a known interfering substance as listed in the rp500 operator's manual.The sensor had been exposed numerous times throughout the use of this cartridge causing intermittent na calibration errors.The source of the contamination should be investigated and eliminated.There were no system or sample errors occurring on instrument (b)(4) on (b)(6) when the suspect samples were run.There were differences in recovery on na, k, cl, thb and lactate suggesting pre-analytical factors.
|