• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - SWINFORD ENTERPRISE TSUNAGU SET CAP KIT; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SWINFORD ENTERPRISE TSUNAGU SET CAP KIT; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number JPC4222
Device Problem Misconnection (1399)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a disconnect kit had been connected with no error, but there was a small gap under the cap.As a result, the next bag change was difficult, and the patient found it difficult to set the uv flash machine.The patient tried another disconnect kit from a different lot with no issues.There was no patient injury or medical intervention indicated at the time of the report.No additional information is available.
 
Manufacturer Narrative
(b)(4).The actual device was not available; however, a companion sample was received for evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed with no issues noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERPRISE TSUNAGU SET CAP KIT
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE - SWINFORD
foxford road
swinford
EI 
Manufacturer (Section G)
BAXTER HEALTHCARE - SWINFORD
foxford road
swinford
EI  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5704483
MDR Text Key46761964
Report Number1416980-2016-10386
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2020
Device Catalogue NumberJPC4222
Device Lot Number15F22H70
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/22/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-