BAXTER HEALTHCARE - SWINFORD ENTERPRISE TSUNAGU SET CAP KIT; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number JPC4222 |
Device Problem
Misconnection (1399)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/25/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Should additional information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a disconnect kit had been connected with no error, but there was a small gap under the cap.As a result, the next bag change was difficult, and the patient found it difficult to set the uv flash machine.The patient tried another disconnect kit from a different lot with no issues.There was no patient injury or medical intervention indicated at the time of the report.No additional information is available.
|
|
Manufacturer Narrative
|
(b)(4).The actual device was not available; however, a companion sample was received for evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed with no issues noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|