• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON MICRO MIST NEBULIZER W/TEE; NEBULIZER ADAPTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON MICRO MIST NEBULIZER W/TEE; NEBULIZER ADAPTOR Back to Search Results
Catalog Number 1884
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Potential lot numbers reported: 74a1602818, 74a1603215, 74b1602277 and 74c1600209.A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.No photo for review.The device history record of (b)(4) that belong to catalog number 1884 has been reviewed and no issues or discrepancies were found which could potentially be related to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved in this complaint.However material from the production line was inspected and no issues were detected that can lead this customer complaint.If device sample becomes available at a later date this complaint will be re-opened.
 
Event Description
The customer alleges the nebulizer kit separated at the universal tubing end of the tubing.It popped off when attached to the flow meter.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the connector was separated from the tubing.It was also found that the tubing contained pip residue.A dimensional inspection was performed and the internal diameter of the universal connector and the external diameter of tubing were found to be within specification.Based on the investigation performed, the reported complaint was confirmed.The disconnection of the connector and the tubing could be caused by an incorrect assembly of the tubing and the connector; therefore , all personnel from the production line were re-trained on the work instruction.In addition, a non-conformance was generated.
 
Event Description
The customer alleges the nebulizer kit separated at the universal tubing end of the tubing.It popped off when attached to the flow meter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON MICRO MIST NEBULIZER W/TEE
Type of Device
NEBULIZER ADAPTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5706537
MDR Text Key46809794
Report Number3004365956-2016-00233
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Respiratory Therapist
Type of Report Initial,Followup
Report Date 05/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date03/02/2021
Device Catalogue Number1884
Device Lot Number74C1600209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-