MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-723LNAS
Device Problems Break (1069); Component Falling (1105); Malposition of Device (2616); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/20/2016
Event Type  malfunction  
Manufacturer Narrative
Unit had minor scratched lcd window, cracked lcd window, cracked case at display window corner, cracked battery tube threads, a test reservoir was installed and did not lock in place due to complete broken reservoir tube lip and missing o-ring (reservoir tube).No belt clip assembly received with pump.Was unable to verify belt clip anomaly.
 
Event Description
Customer reported via phone call to have physical damage of insulin pump.Customer reported that reservoir compartment was broken and reservoir could not stay in place.Customer does not know how damage occurred, but may have been dropped.Customer's blood glucose was 115 mg/dl.Customer was advised that insulin pump would need to be replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5710357
MDR Text Key47227433
Report Number3004209178-2016-52774
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 05/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723LNAS
Device Catalogue NumberMMT-723LNAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received05/26/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age66 YR
-
-