Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the battery impedance trend was rising.Recommended replacement time (rrt) was triggered on (b)(6) 2016 due to impedance.Daily battery trend data shows gradual rise of battery impedance.Early rrt alert, without corresponding battery depletion.This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.
|