• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEE BIPLANE; SYSTEM, XRAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEE BIPLANE; SYSTEM, XRAY, ANGIOGRAPHIC Back to Search Results
Model Number 10094141
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/17/2016
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that the artis zee biplane system went down during a tavr procedure.The patient was safely removed from the system and transferred to an alternative system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The non-conformity leading to the blockage of x-ray imaging was the result of an interrupted coolant flow when the fd cooling unit coupling failed.The couplings to connect the cooling circuit from the system to the system cabinet (incl.Fd cooling unit) are made of plastic.During exchange of the fd cooling unit the coupling has to be loosened by pressing the latch.As the plug and the mechanical structure around the latch are made of plastic the frame holding the latch can break when the applied force is too high.Siemens has initiated a field corrective action for this issue which has been reported to the fda via 806.10 report # 2240869-02/19/16-0006-c.This corrective action eliminates the root cause of the problem and prevents the possibility of reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE BIPLANE
Type of Device
SYSTEM, XRAY, ANGIOGRAPHIC
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key5717099
MDR Text Key48153930
Report Number2240869-2016-43050
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K073290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 06/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Technologist
Device Model Number10094141
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/06/2017
Date Manufacturer Received12/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-