Model Number 4415
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/23/2012
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported complaint was confirmed.The field service representative (fsr) was able to confirm the complaint and determined valve 5 was buzzing.The fsr removed and cleaned the valve.The unit operated to manufacturer specifications and was returned to clinical use.An mdr remediation activity related to manufactured devices with a fda product code ¿dwc: controller, temperature, cardiopulmonary bypass heater cooler" was conducted.This report is a result of the ¿heater ¿ cooler response remediation protocol 02-jun-2016.¿.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the cooler heater valve was chattering.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TCM II HEATING AND COOLING SYSTEM
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5720250
MDR Text Key47239352
Report Number1828100-2016-00421
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 06/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received06/02/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-