TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TCM II HEATING AND COOLING SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
|
Back to Search Results |
|
Model Number 4415 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/23/2012 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.The field service representative (fsr) was able to confirm the complaint and determined valve 5 was buzzing.The fsr removed and cleaned the valve.The unit operated to manufacturer specifications and was returned to clinical use.An mdr remediation activity related to manufactured devices with a fda product code ¿dwc: controller, temperature, cardiopulmonary bypass heater cooler" was conducted.This report is a result of the ¿heater ¿ cooler response remediation protocol 02-jun-2016.¿.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the cooler heater valve was chattering.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|