• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TCM II HEATING AND COOLING SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TCM II HEATING AND COOLING SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4415
Device Problems Corroded (1131); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/17/2012
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported complaint was confirmed.The user facility¿s biomedical engineer (biomed) troubleshot the issue to solenoid valve 5.This valve was intermittently making a noise.The biomed removed the valve and cleaned it out.It was then placed back into the unit.The biomed was concerned it was still making some noise and tried to order a new valve.The field service representative (fsr) was unable to duplicate the noise.The fsr replaced the valve per customer request.As the biomed cleaned the valve and found the noise had improved.Visual inspection of the returned valve by the investigator determined corrosion is present inside the valve along the diaphragm.A mdr remediation activity related to manufactured devices with a fda product code ¿dwc: controller, temperature, cardiopulmonary bypass heater cooler" was conducted.This report is a result of the ¿heater ¿ cooler response remediation protocol 02-jun-2016.¿.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the cooler heater unit was noisy.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TCM II HEATING AND COOLING SYSTEM
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5720360
MDR Text Key47234161
Report Number1828100-2016-00424
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2012
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/02/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-