TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TCM II HEATING AND COOLING SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
|
Back to Search Results |
|
Model Number 4415 |
Device Problems
Device Contamination with Chemical or Other Material (2944); Noise, Audible (3273)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/18/2013 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.The deficiency of the loud buzzing noise was confirmed by the user facility¿s biomedical engineer (biomed).The issue did not cause a functional failure.The biomed cleaned/descaled unit.The noise was not present after the cleaning.User facility did not request service on the unit.A mdr remediation activity related to manufactured devices with a fda product code ¿dwc: controller, temperature, cardiopulmonary bypass heater cooler" was conducted.This report is a result of the ¿heater ¿ cooler response remediation protocol 02-jun-2016.¿.
|
|
Event Description
|
It was reported that during the use of the device for a non-clinical activity, the perfusionist (ccp) discovered that the cooler heater unit was making a loud buzzing noise.There was no patient involvement.
|
|