• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TCM II HEATING AND COOLING SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TCM II HEATING AND COOLING SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4415
Device Problems Device Contamination with Chemical or Other Material (2944); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Date 07/18/2013
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported complaint was confirmed.The deficiency of the loud buzzing noise was confirmed by the user facility¿s biomedical engineer (biomed).The issue did not cause a functional failure.The biomed cleaned/descaled unit.The noise was not present after the cleaning.User facility did not request service on the unit.A mdr remediation activity related to manufactured devices with a fda product code ¿dwc: controller, temperature, cardiopulmonary bypass heater cooler" was conducted.This report is a result of the ¿heater ¿ cooler response remediation protocol 02-jun-2016.¿.
 
Event Description
It was reported that during the use of the device for a non-clinical activity, the perfusionist (ccp) discovered that the cooler heater unit was making a loud buzzing noise.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TCM II HEATING AND COOLING SYSTEM
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48102
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5720589
MDR Text Key48158002
Report Number1828100-2016-00428
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 06/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/02/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-