No further investigation is needed since the device works within specification.The customer reported his dissatisfaction with the new version of hl20 regarding the bubble alarm.It has to be noted, that the older version of hl20 device as well as the 3rd edition device do not generate an alarm, if the activation of the bubble sensor is removed.But of course in both cases the pump will stop when bubble sensor and level sensor are activated.Therefore, no product malfunction occurred.So the hl20 has worked within specification.Furthermore, the medical staff in (b)(6) is trained into the functionality of hl devices otherwise the person is not allowed to work on this device.Thus the failure could not be confirmed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition no device malfunction has been determined.No corrective action is needed.
|