• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG MAQUET HCU40 DEVICE; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG MAQUET HCU40 DEVICE; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701044054
Device Problem Contamination (1120)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/10/2016
Event Type  malfunction  
Manufacturer Narrative
On 2015-11-30, the fsca 2015-11-30 was issued regarding revised decontamination procedures for heater and heater-cooler systems from maquet.This fsca was put on hold a few days later with fsca 2015-12-10.A new fsca will be issued as soon as the new disinfection procedure is properly validated and can be launched.The new ifu is currently under revision and will contain the instructions for the new disinfection procedure.Q4 is scheduled for the market launch of this new disinfection procedure.
 
Event Description
It was reported that the water of a hcu40 was contaminated with "cupriavidus pauculus".No patient involvement was reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU40 DEVICE
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key5723574
MDR Text Key47893965
Report Number8010762-2016-00385
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701044054
Device Catalogue Number701044054
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/10/2016
Date Device Manufactured12/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-