Model Number 9438-06
Device Problem Loss of Data (2903)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/18/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Distributor contacted dexcom on behalf of the patient on (b)(6) 2016, to report gaps in data that occured on (b)(6) 2016.There was no report of injury or medical intervention.No additional event or patient information is available.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Voltage testing was performed and resulted in 0 voltage.A review of the data log confirmed the reported event of gaps in error.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5731323
MDR Text Key47588997
Report Number3004753838-2016-21821
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9438-06
Device Catalogue NumberSTT-GF-004
Device Lot Number5206805
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-