Model Number N/A
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Type  Injury  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: under precautions, "intraoperative fracture or breaking of instruments has been reported for general instruments.".
 
Event Description
During an initial hip arthroplasty the thread on the tip of the inserter fractured.The fractured piece was removed from the patient.
 
Manufacturer Narrative
Examination of returned device found no evidence of product non-conformance.Review of the device confirmed the reported condition.A conclusive root cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
HIP, PROSTHESIS
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key5732535
MDR Text Key47624744
Report Number0001825034-2016-02122
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number888180
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Device Age1 YR
Date Manufacturer Received06/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberNI
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-