• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. IMPACTOR-POLY LINER-36MM; HIP PROSTHESIS, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER, + ADDITIVE, POROUS, U

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. IMPACTOR-POLY LINER-36MM; HIP PROSTHESIS, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER, + ADDITIVE, POROUS, U Back to Search Results
Catalog Number 112636
Device Problems Break (1069); Material Deformation (2976); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/20/2016
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The 32 and 36 liner impactor heads both were damaged and spun on the liner impactor shaft.We finished the case without delay of adverse affect.
 
Manufacturer Narrative
The event is regarding damaged threads involving a mako impactor was reported.The event was confirmed.Method & results: -device evaluation and results: damage was observed on the threads of the impactor consistent with cross threading with the mating instrument and most likely from field use.-medical records received and evaluation: no patient medical records were available for review.-device history review: review of the device history records indicate devices were manufactured and accepted into final stock on with no reported discrepancies.-complaint history review: there has been 1 other event for the lot referenced.Conclusions: the investigation confirmed that the device has damaged threads from field use.There is no indication at this time that the design, materials, or manufacturing of the subject device contributed to the event.If additional information becomes available, this investigation will be reopened.
 
Event Description
The 32 and 36 liner impactor heads both were damaged and spun on the liner impactor shaft.We finished the case without delay of adverse affect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPACTOR-POLY LINER-36MM
Type of Device
HIP PROSTHESIS, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER, + ADDITIVE, POROUS, U
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5732618
MDR Text Key47642609
Report Number3005985723-2016-00189
Device Sequence Number1
Product Code OQG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number112636
Device Lot Number120080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/30/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-