Model Number 3100A
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/18/2016
Event Type  Injury  
Manufacturer Narrative
Should any additional information be received by the customer then an additional report will be submitted.(b)(4).Device evaluation: the reporter of the incident stated that it is was discovered through an internal investigation that the nurse bumped into this unit causing a disconnect of the device, ultimately resulting in the dumping of pressure and the oscillator stopping.A carefusion field service representative went onsite to the reporter's facility and evaluated the unit.The unit passed inspection, meeting manufacturer specifications.There was no failure detected.
 
Event Description
The customer reported that this 3100a high frequency oscillating ventilator was being used on a patient when it unexpectedly stopped oscillating.The customer stated that alternate ventilation was provided by placing the patient on a conventional ventilator and there was no patient injury or harm associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
7149227830
MDR Report Key5740289
MDR Text Key47892772
Report Number2021710-2016-03920
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/24/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-