Model Number M-5723-17
Device Problem Device Operates Differently Than Expected (2913)
Patient Problems Cardiac Arrest (1762); Cardiac Tamponade (2226)
Event Date 05/30/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation procedure with a soundstar® eco diagnostic ultrasound catheter and suffered a cardiac tamponade and cardiac arrest which required pericardiocentesis and surgical intervention.When the soundstar eco catheter was connected, a non mdr reportable magnetic sensor error occurred.The issue was resolved by changing the catheter to another one.A soundstar merge was then done and a transseptal puncture was performed.The replacement soundstar eco catheter was then inserted into the right ventricle (rv) for monitoring effusions.Right after the pulmonary vein isolation (pvi) was started, the patient¿s blood pressure dropped and a cardiac tamponade was confirmed.A pericardiocentesis was conducted and an unknown amount of fluid was removed.The patient became stable after a few hours and the pvi was re-conducted but the patient¿s blood pressure dropped again.Another pericardiocentesis was performed which was not successful and the patient was taken for surgery.Cardiac arrest occurred after surgery and at that time the patient was unstable.The patient did require hospitalization due to the condition worsening.The patient has since fully recovered.The physician¿s opinion regarding the cause of this adverse event is that this is procedure related.The physician commented the adverse event might have been caused by the transseptal needle (brand unknown) during the transseptal phase or by the second soundstar eco catheter that was inserted into the rv.Generator settings include: power control mode.There were no error messages observed on biosense webster equipment during the procedure.This event is usually reported under the ablation catheter, however since the physician believes this catheter was the possible causative device it is only being reported under this soundstar catheter.
 
Manufacturer Narrative
Additional information was received on this product on 06/27/2016.The manufacturing date is 02/11/2016; the expiration date is 02/28/2017.Concomitant product information received 06/29/2016: webster cs with auto id ((b)(4)) model# d-1353-03-s, lot# 17127539m.The bwi failure analysis lab received the device for evaluation on 07/05/2016.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.Manufacturer's ref.No: (b)(4).
 
Manufacturer Narrative
(b)(4).It was reported that a patient underwent an atrial fibrillation procedure with a soundstar® eco diagnostic ultrasound catheter and suffered a cardiac tamponade and cardiac arrest which required pericardiocentesis and surgical intervention.When the soundstar eco catheter was connected a non mdr reportable magnetic sensor error occurred.Upon receipt, the catheter was visually inspected and it was found in normal conditions.A carto test and coil disconnection test was performed and catheter failed specifications.The catheter was dissected and it was found that the y sensor coil did not pass specifications.Investigation found y sensor was discontinued at the tip.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Based on available analysis results, there was a coil sensor failure related to the recognition between the catheter and the carto 3 system.The failure of the coil sensor did not contribute to the tamponade.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOUNDSTAR® ECO DIAGNOSTIC ULTRASOUND CATHETER
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BIOSENSE WEBSTER, INC. (IRWINDALE)
15715 arrow highway
irwindale CA 91706
Manufacturer (Section G)
BIOSENSE WEBSTER, INC. (IRWINDALE)
15715 arrow highway
irwindale CA 91706
Manufacturer Contact
joaquin kurz
33 technology drive
irvine, CA 92618
9497893837
MDR Report Key5743625
MDR Text Key48003895
Report Number2029046-2016-00073
Device Sequence Number1
Product Code OBJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K140318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Model NumberM-5723-17
Device Catalogue Number10439011
Device Lot NumberOEM_M-5723-17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-