SYNTHES HAGENDORF APPLICATION INSTRUMENT FOR STERNAL ZIPFIX; CERCLAGE FIXATION Back to Search Results
Catalog Number 03.501.080
Device Problems Sticking (1597); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Device is an instrument and is not implanted/explanted.Lot number provided (873513) was not able to be verified in the system.Without a lot number the device history records review could not be completed.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported the trigger for the application instrument for sternal zipfix would not release; the trigger seemed to be sticking.During the patient's initial surgery on (b)(6) 2016, the tensioner would not tighten zip fix ties as desired.When checked after surgery, the tensioner trigger was sticking.There was no surgical delay or patient harm.The procedure was successfully completed.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
No service history review can be performed as part number 03.501.080 with lot number(s) 8733513 is a lot/batch controlled item.The manufacture date of this item is january 10, 2014.The source of the manufacture date is the release to warehouse date.The service history review is unconfirmed.Manufacturing location: (b)(4).Manufacturing date: january 06, 2014.No non-conformance reports were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.A service and repair evaluation was completed: the customer reported the trigger would not release.The repair technician reported lube/oil/clean as the reason for repair.The cause of the issue is unknown.No parts were replaced.The item was repaired per the inspection sheet, passed synthes final inspection and will be returned to the customer upon completion of the service and repair process.The evaluation was confirmed.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLICATION INSTRUMENT FOR STERNAL ZIPFIX
Type of Device
CERCLAGE FIXATION
Manufacturer (Section D)
SYNTHES HAGENDORF
im bifang 6
hagendorf PA CH461 4
SZ  CH4614
Manufacturer (Section G)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH461 4
SZ   CH4614
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5745820
MDR Text Key48099513
Report Number2520274-2016-13274
Device Sequence Number1
Product Code JDQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK110789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.501.080
Device Lot Number8733513
Other Device ID Number(01)10887587010847(10)8733513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age82 YR
Patient Weight98
-
-