• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC THINPREP 5000 PROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC THINPREP 5000 PROCESSOR Back to Search Results
Device Problems Human Factors Issue (2948); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  No Answer Provided  
Event Description
A customer in (b)(6) first reported an incident in (b)(6) 2016 to (b)(6) field service engineer (fse) where their thinprep 5000 processor with autoloader etched a slide with a wrong id number.*due to a misunderstanding of the event between the customer and (b)(6) field service engineer (fse), initially the incident was not considered a reportable event by the fse.The fse understood the event as the instrument could not read a slide id of one particular slide.The customer actually reported that the autoloader etched a slide with a wrong id number.The misunderstanding was discovered on (b)(6) when the customer reviewed the field service report from the fse visit in (b)(6) and noticed the miscommunication of the event, then notified hologic customer service and this event was elevated as a pae.Hologic's field service engineer (fse) confirmed with the customer that the patients sample was found and reprocessed.The barcode produced was correct and matched the vial id upon processing the sample the second time.Therefore, no patient recall was necessary.In addition, it was confirmed by the fse that no error code was presented to the operator.Fse confirmed but unable to reproduce the error, one slide printed with wrong id.Fse found the following to be the most likely cause of the error: vial barcode reader.Performed tsb 00882 to resolve the problem (see sr #(b)(4)).Ran go no-go test to verify instrument operation.Processed samples to confirm operation.Instrument operational.Although no patients needed to be recalled as a result of this incident this is a reportable event since the instrument did not perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THINPREP 5000 PROCESSOR
Type of Device
THINPREP 5000 PROCESSOR
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer Contact
eva maxwell
250 campus drive
marlborough, MA 01752
MDR Report Key5747724
MDR Text Key49300830
Report Number1222780-2016-00163
Device Sequence Number1
Product Code MKQ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 06/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received06/02/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-