The customer reported that the companion 2 driver compressors were cycling on and off.The customer also reported that the patient was subsequently switched to the backup companion 2 driver.There was no reported adverse patient impact.The companion 2 driver was returned to syncardia for investigation.The patient file was reviewed and revealed numerous external air connected notifications.This is an indication that the necessary conditions required for the driver to operate exclusively on external air were not consistently met, thereby causing the driver to disable the external air connected icon and activate the primary compressor.This confirmed the customer-reported compressor cycling issue.The failure investigation confirmed that the manual pressure regulator could not maintain the proper set point; therefore, the manual pressure regulator was taken out of service and replaced.The root cause for this failure mode is most likely the inability of the manual pressure regulatory to maintain proper set point.The issue of compressor cycling because of a malfunction of the manual pressure regulator is being investigated in a capa (corrective or preventive action).The capa will document the investigation including, but not limited to, potential root cause and corrective actions associated with compressor cycling because of a malfunction of the manual pressure regulator.The companion 2 driver was serviced and passed all functional and performance testing before being placed into finished goods.Although the compressors were cycling on and off, the risk to the patient would have been low because the driver continued to perform its life-sustaining functions.Additionally, centers are equipped with backup drivers.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|