• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 723585
Device Problem Charred (1086)
Patient Problem Not Applicable (3189)
Event Date 06/08/2016
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) evaluated the instrument.The fse found that the customer had connected the instrument plug that was defective and not compatible with the instrument specifications.This caused the cable plug to be burnt.The fse replaced the damaged cable.The repairs were verified by the fse.(b)(4).
 
Event Description
The customer reported the plug of the coulter lh 780 hematology analyzer was burnt when connected to an outlet.There was smoke reported by the customer.The fire department was not called and there was no death or injury as a result.There was no medical attention required by any operator.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 780 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-s08
miami, FL 33196
3053802031
MDR Report Key5752095
MDR Text Key48303112
Report Number1061932-2016-00574
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K061616
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number723585
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-