• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACUMED LLC ARH SLIDE-LOC¿ HEAD 20MM, LEFT; PROSTHESIS, ELBOW, HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACUMED LLC ARH SLIDE-LOC¿ HEAD 20MM, LEFT; PROSTHESIS, ELBOW, HEAD Back to Search Results
Model Number 5001-0220L-S
Device Problems Detachment Of Device Component (1104); Failure to Align (2522)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
Additional mdrs associated with this event: mdr 3025141-2016-00137 neck, mdr 3025141-2016-00138 stem.
 
Event Description
Arh slide-loc product was implanted on (b)(6) 2016.On (b)(6) 2016, during a routine follow up examination, it was discovered that the head/neck assembly had disassociated from the stem.The implants were explanted on (b)(6) 2016.The sales rep reports that the laser lines on the head and neck were not aligned after rotation to lock; this leads to incomplete locking of the implants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARH SLIDE-LOC¿ HEAD 20MM, LEFT
Type of Device
PROSTHESIS, ELBOW, HEAD
Manufacturer (Section D)
ACUMED LLC
5885 nw cornelius pass road
hillsboro OR 97124 9432
Manufacturer (Section G)
ACUMED LLC
5885 nw cornelius pass road
hillsboro OR 97124 9432
Manufacturer Contact
micki lehman
5885 nw cornelius pass road
hillsboro, OR 97124-9432
5036279957
MDR Report Key5757100
MDR Text Key48400075
Report Number3025141-2016-00136
Device Sequence Number1
Product Code KWI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131845
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 05/31/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date12/20/2022
Device Model Number5001-0220L-S
Device Catalogue Number5001-0220L-S
Device Lot Number359415
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/31/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-