The cause of the shifted high aptt results is unknown.Siemens evaluated the data.The shift high was observed in some patients, not all patients.Actual patient values and repeat testing was requested, however, was not provided by the customer to determine if values were truly erroneous.They observed a shift upward in some patient recovery where they expect normal values since (b)(6) 2016.They did not observe a shift upward in control recovery.The issue resolved with a new shipment of the lot of reagent that had been in use at the account since (b)(6) 2015.The account believes this was resolved with a new shipment of reagent.No further evaluation of the devices is required.
|