The cause of the falsely elevated aptt results is unknown.The true recovery of the patient samples is not clear with the information provided.Siemens is unable to rule out improper collection or handling of the samples.Per siemens healthcare support center analysis, the original sample recovery of 62 seconds produced a recovery of 33.8 seconds when repeat analysis was performed on the same sample after 5 hours.The sample, when aged 5 hours was no longer within acceptable stability for the aptt assay when assessing for unfractionated heparin.Samples collected for coagulation testing for patients on unfractionated heparin therapy must be centrifuged within 1 hour of collection and tested within 4 hours.Delay in centrifugation will lead to platelet neutralization of the heparin in the plasma.If the sample was left in the original tube between the original run and repeat analysis, the platelets may re-suspend in the plasma and also lead to platelet neutralization of the heparin in the plasma.Neutralization of heparin in the plasma would lead to falsely decreased values.The account stated that there may have been a bubble aspirated with the evening sample.The instrument and reagent are performing within specifications per qc results.There were no report of discordant or questionable results on any other patients.No further evaluation of the devices is required.
|