Model Number 3788
Device Problems Nonstandard Device (1420); Use of Device Problem (1670)
Patient Problem Inadequate Pain Relief (2388)
Event Date 06/03/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).This ipg serial number was included in a field advisory.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient did not charge her scs ipg as recommended.Consequently, the ipg battery depleted and the ipg was explanted and replaced with a new one.Stimulation therapy was reportedly restored postoperative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5765104
MDR Text Key48646317
Report Number1627487-2016-03401
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 06/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/28/2013
Device Model Number3788
Device Lot Number3486380
Was Device Available for Evaluation? No
Date Manufacturer Received06/06/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-002-R
Patient Sequence Number1
Treatment
MODEL 3186 (X2), SCS LEAD
Patient Outcome(s) Other;
Patient Age59 YR
-
-