Model Number P3200K000015
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/21/2016
Event Type  malfunction  
Manufacturer Narrative
The account requested a hill-rom technician repair the bed.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Examine and test the communication junction box.Make sure the sidecom® communication system feature works correctly.Examine the communication cable, include the male and female pins in the plug.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.No further information is available on the repair of the bed at this time.The investigation is ongoing, however if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.
 
Event Description
Hill-rom received a report from the account stating the nurse call would not work.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key5775088
MDR Text Key49675706
Report Number1824206-2016-00289
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Hospital Service Technician
Type of Report Initial
Report Date 06/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP3200K000015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/21/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-