• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. 36" EVOLUTION STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. 36" EVOLUTION STERILIZER Back to Search Results
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2016
Event Type  No Answer Provided  
Manufacturer Narrative
A steris service technician inspected the unit and identified that the door, and door connection points on the unit, had become corroded.During the evaluation of the unit, the technician determined that steam was entering the door assembly resulting in damage to the components that secure the door to the sterilizer.The sterilizer is 34 years old and is not maintained by steris.The root cause of the reported event is improper maintenance of the unit which caused the reported damage and corrosion of the door assembly.The user facility reported the sterilizer will not be replaced as the facility is closing and will move to a new location.
 
Event Description
The user facility reported that the door of their 36" evolution sterilizer came off during a processing cycle.No injury, procedural delay or cancellation was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36" EVOLUTION STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX   67190
Manufacturer Contact
kathryn cadorette
5960 heisley road
mentor, OH 44060
4403927231
MDR Report Key5776065
MDR Text Key49766906
Report Number3005899764-2016-00049
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-