A steris service technician inspected the unit and identified that the door, and door connection points on the unit, had become corroded.During the evaluation of the unit, the technician determined that steam was entering the door assembly resulting in damage to the components that secure the door to the sterilizer.The sterilizer is 34 years old and is not maintained by steris.The root cause of the reported event is improper maintenance of the unit which caused the reported damage and corrosion of the door assembly.The user facility reported the sterilizer will not be replaced as the facility is closing and will move to a new location.
|