• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB CRYSTALENS ACCOMMODATING IOL; LENS, INTRAOCULAR, ACCOMMODATIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB CRYSTALENS ACCOMMODATING IOL; LENS, INTRAOCULAR, ACCOMMODATIVE Back to Search Results
Model Number AT50AO
Device Problem Malposition of Device (2616)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/09/2016
Event Type  Injury  
Manufacturer Narrative
Investigation of this event is in progress.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was reported that the lens in the patient's right eye was malpositioned.At one and a half months post-op, the doctor was unable to reposition it so the lens was explanted and the same model and diopter was implanted.
 
Manufacturer Narrative
The lens was returned to bausch and lomb.Visual inspection found the lens in three pieces.The optic has been torn in half.One plate is torn off and one haptic is bent.The cause of the damage cannot be determined.Functional and dimensional testing cannot be performed due to the damage.The device history record was reviewed and there were no discrepancies or unusual finding that relate to the reported issue.Based on the current information the root cause of the event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYSTALENS ACCOMMODATING IOL
Type of Device
LENS, INTRAOCULAR, ACCOMMODATIVE
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
10574 acacia street suite d1
rancho cucamonga CA 91730
Manufacturer Contact
sharon spencer
50 technology drive west
irvine, CA 92618
9493985698
MDR Report Key5777540
MDR Text Key49020454
Report Number0001313525-2016-00303
Device Sequence Number1
Product Code NAA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Model NumberAT50AO
Device Lot Number7368716
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age35 YR
-
-