MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 2088TC/52
Device Problems Device Dislodged or Dislocated (2923); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/25/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the lead dislodged 3 days post implant, on (b)(6) 2016 the lead was repositioned successfully.A couple of days later the lead dislodged once more, the physician decided to reposition the lead, during procedure he noted a white element that could be the reason why the lead lost fixation.The lead was explanted and replaced successfully, the patient's condition was good post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key5778625
MDR Text Key49056275
Report Number2017865-2016-04641
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Model Number2088TC/52
Device Lot NumberP000024641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-