SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEEGO; INTERVENTIONAL, FLUOROSCOPIC, XRAY
|
Back to Search Results |
|
Model Number 10280959 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/14/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.(b)(6).
|
|
Event Description
|
It was reported to siemens that a malfunction occurred on the artis zeego system.During a procedure, x-ray was not possible and was stopped by angiomatic.As a result, the examination was aborted.We are unaware of any impact to the state of health of the patient involved.
|
|
Manufacturer Narrative
|
Siemens has completed an investigation of the reported event.The root cause was determined to be a failure of the d601 component due to improper connection or improper cabling of the component to the system.At the system itself, no additional failures could be identified,.The d601 component was replaced and the problem did not reoccur.There is no negative awareness in regards to the quality or performance of the affected component, especially in regards to the connection of the component to the system, therefore, no further actions will be taken.
|
|