Model Number N/A |
Device Problems
Component Falling (1105); Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This user facility is outside of the united states.Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.
|
|
Event Description
|
During a procedure, a pin on the cutting guide fell off.No patient injury or delay was reported as a result of the event.
|
|
Manufacturer Narrative
|
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.
|
|
Event Description
|
During a procedure, a pin on the cutting guide fell off into the patient's wound.The piece was retrieved and nothing remained in the patient.There was a delay of approximately 15 minutes as a result of the event.
|
|
Manufacturer Narrative
|
This follow up report is being filed to relay product evaluation results.Review of manufacturing history found no evidence of product nonconformance.Dimensional analysis performed revealed the device conformed to specifications.Examination of the returned device revealed a tarnished appearance and excessive damage, evidence of extensive use and multiple sterilizations.The spring and plug had no evidence of residue.A conclusive root cause of the event cannot be determined with the available information.
|
|
Search Alerts/Recalls
|