• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXF MP SLOT POST CUT GD MED; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF MP SLOT POST CUT GD MED; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problems Component Falling (1105); Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2016
Event Type  malfunction  
Manufacturer Narrative
This user facility is outside of the united states.Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.
 
Event Description
During a procedure, a pin on the cutting guide fell off.No patient injury or delay was reported as a result of the event.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.
 
Event Description
During a procedure, a pin on the cutting guide fell off into the patient's wound.The piece was retrieved and nothing remained in the patient.There was a delay of approximately 15 minutes as a result of the event.
 
Manufacturer Narrative
This follow up report is being filed to relay product evaluation results.Review of manufacturing history found no evidence of product nonconformance.Dimensional analysis performed revealed the device conformed to specifications.Examination of the returned device revealed a tarnished appearance and excessive damage, evidence of extensive use and multiple sterilizations.The spring and plug had no evidence of residue.A conclusive root cause of the event cannot be determined with the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF MP SLOT POST CUT GD MED
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
daniel tilbury
waterton industrial estates
bridgend 
0441656655
MDR Report Key5780529
MDR Text Key49131152
Report Number3002806535-2016-00550
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
PP010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 08/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number32-422981
Device Lot NumberZB140602
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/10/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2014
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
-
-