SMITH & NEPHEW, INC. ARTHROSCOPE ACLV 2.7X30D SHORT Back to Search Results
Catalog Number 7205682
Device Problem Device Or Device Fragments Location Unknown (2590)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 05/25/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
User facility medwatch report received indicates that when the orthopedic hand surgeon took the scope out and put it back into the patient's wrist, he could not see very well.Scope was sent out for evaluation and repair.Facility biomed examined the scope and noted that the tiny glass/clear cover that goes over the lens, was missing.Md was notified and disclosure was done with the patient.Md to continue to follow up with the patient.Md determined the piece is so small it would not be able to be located/noted on x-ray and more than likely would have been washed out of the patient during copious irrigation that took place during the procedure.There have been no complaints of pain/issues from the patient.
 
Manufacturer Narrative
Visual inspection and functional testing could not be performed because the device in question will not be returned for evaluation.Thus, the complaint could not be verified and a root cause could not be determined with confidence.It is difficult to perform an accurate evaluation of a failure without having the failed product to look at.As such the complaint is being closed without conclusion.However, if the product is returned in the future the complaint can be reopened and evaluated.Our quality department will continue to monitor for trends.No further investigation is required.Evaluation codes updated to indicate that manufacturing review was performed.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHROSCOPE ACLV 2.7X30D SHORT
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key5780589
MDR Text Key49126193
Report Number3003604053-2016-00031
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K962075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number7205682
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-