It was reported that a foreign material was noted at the tip of the guidewire.A 180 x 25cm fathom - 16 was used during a y-90 procedure.During preparation, the guidewire was unable to be inserted into an unknown microcatheter.After several tries, it was noted that the tip of the guidewire had a ball of extra material.The physician tried to remove the extra material using his fingers.After several attempts, the extra material was removed.The procedure was completed with another of the same device.No patient complications were reported.
|