MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ZM-521PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ZM-521PA; TRANSMITTER Back to Search Results
Model Number ZM-531PA
Device Problems Overheating of Device (1437); Battery Problem (2885); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2016
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the transmitter became very hot.They tried replacing the batteries but the problem still remains.The customer reports there was no fluid intrusion or physical damage.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that the transmitter became very hot.They tried replacing the batteries but the problem still remains.The customer reports there was no fluid intrusion or physical damage.
 
Manufacturer Narrative
Manufacturer narrative: the customer reported that the transmitter became very hot.They tried replacing the batteries but the problem still remains.The customer reports there was no fluid intrusion or physical damage.The unit was evaluated.The batteries required for this investigation were not returned.New batteries were inserted into the unit and heating could not be duplicated.Inspection of the negative contacts shows resin melting at the spring, which per an nkc investigation on a similar incident, is indicative of improper battery insertion.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-521PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key5793138
MDR Text Key50179230
Report Number8030229-2016-00314
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/13/2016,06/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-531PA
Device Catalogue NumberZM-531PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/13/2016
Distributor Facility Aware Date06/13/2016
Device Age15 MO
Event Location Hospital
Date Report to Manufacturer07/13/2016
Date Manufacturer Received07/13/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/05/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-