It was reported that a surgeon was seeing high impedance during a patient's initial implant surgery.High impedance was found during multiple system diagnostic tests.The surgeon then tried performing generator diagnostics and found the results were within normal limits.A replacement lead was used and the system diagnostics showed results within normal limits.The lead that was attempted to be used in surgery was returned in one piece and analysis was approved on 07/12/2016.The condition of the returned lead assembly was consistent with conditions that typically exist following an attempted implant procedure.Continuity checks of the returned lead assembly were performed with no discontinuities identified.Overall length and resistance measurements of the complete returned lead were determined to be in compliance with those defined in the manufacturing specifications.Five (5) half sets of screw marks were observed near the end of the connector pin, indicating that the lead pin had not been fully inserted into the generator block.The lead was inserted into a test generator and verified that there were no abnormalities with the connector boot.Based on the findings in the product analysis lab, there is no evidence to suggest a discontinuity in the returned portion of the device which may have contributed to the reported high impedance.Incomplete insertion of the connector pin was the most likely cause for the observed high impedance condition during the implant of this lead.A review of the generator's device history record showed that the generator had passed all quality inspections prior to release for distribution.
|
Describe event or problem, corrected data: the initial report inadvertently did not include the intraoperative impedance values.Relevant tests/laboratory data, including dates, corrected data: the initial report inadvertently did not include the intraoperative impedance values.
|