DEPUY SYNTHES POWER TOOLS BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.710_LL |
Device Problems
Corroded (1131); Loss of Power (1475); Sticking (1597); Device Inoperable (1663); Naturally Worn (2988)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/09/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Gtin is unavailable as the product was made prior to compliance date; (b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition of the silver knob needed to be checked on the device was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had no power, and the trigger was sticking.It was further noted that the electric motor was not functional, corroded, and the trigger components were worn.The assignable root cause was determined to be due to strain/wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported that during pre-surgery testing it was observed that the "silver knob" of the battery oscillator device needed to be checked.During in-house engineering evaluation it was found that the device had no power and did not function, and the trigger was sticking.It was further noted that the electric motor was not functional, corroded, and the trigger components were worn.It was reported that there was no delay in a scheduled surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no patient or user injuries reported.It was reported there was no medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|