DEPUY SYNTHES POWER TOOLS BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.710_LL
Device Problems Corroded (1131); Loss of Power (1475); Sticking (1597); Device Inoperable (1663); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2016
Event Type  malfunction  
Manufacturer Narrative
Gtin is unavailable as the product was made prior to compliance date; (b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition of the silver knob needed to be checked on the device was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had no power, and the trigger was sticking.It was further noted that the electric motor was not functional, corroded, and the trigger components were worn.The assignable root cause was determined to be due to strain/wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported that during pre-surgery testing it was observed that the "silver knob" of the battery oscillator device needed to be checked.During in-house engineering evaluation it was found that the device had no power and did not function, and the trigger was sticking.It was further noted that the electric motor was not functional, corroded, and the trigger components were worn.It was reported that there was no delay in a scheduled surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no patient or user injuries reported.It was reported there was no medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5793821
MDR Text Key50410329
Report Number8030965-2016-14200
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.710_LL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2016
Date Manufacturer Received07/05/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-