This report is being filed to provide additional information.Investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Per terumo bct's internal documentation, sterilization of the product must be within two months before and two months after the date indicated on the product label.Therefore, although the product is labeled as having expired on june 1, 2016, its sterility is maintained through (b)(6) 2016 plus two months, or (b)(6) 2016.Therefore, there was no risk for use of the set at the time that the customer performed the procedure.According to the device history record, this kit went into the sterilization chamber on (b)(6) 2014 and its expiration date is listed as june 1, 2016.The date of the procedure was (b)(6) 2016.This is within the validated sterilization window of two years plus two months.Root cause: the customer did not have a kit available for use at the time of the procedure, and understood that the kit they used was expired.
|
This report is being filed to provide additional information.Per terumo bct's internal documentation, spectra disposable sets stored for 2 years and 2 months and testing results concluded that the aging does not adversely affect the sterility or performance of the disposable nor does it cause any negative cytotoxicity effects.Updated root cause: the customer understood that the kit was expired and due to the customer's inventory management, an overnight shipment had to be requested.This shipment did not arrive in time for the procedure.
|